Company record
Tobii Dynavox Llc
0 FDA warning letters since 2023 and 2 recalls in the last 12 months.
Get an email the day FDA posts anything new about Tobii Dynavox Llc.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Aug 26 | Class II | Tobii Dynavox Llc Device |
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798 Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure. |
| Aug 26 | Class II | Tobii Dynavox Llc Device |
TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799 Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure. |